Every reusable instrument in a dental practice follows the same loop: used, cleaned, checked, packed, sterilised, stored and used again. Dental assistants run this loop dozens of times a day. Done properly, it is invisible to patients. Done badly, it can put them at serious risk. Here is each stage and what good practice looks like.
The standard behind it
Reprocessing in Australian dental practices follows AS 5369, the Australian Standard for reprocessing reusable medical devices, which replaced the older office-based standard AS/NZS 4815. It also follows the Dental Board's infection control guidelines and the manufacturers' instructions for each instrument and machine. Your practice's protocols should reference these documents.
1. Point-of-use and transport
At the chair, remove gross debris and dispose of sharps immediately. Place instruments in a rigid, closed container to carry them to the reprocessing room. Never carry loose instruments by hand.
2. Cleaning
Cleaning removes the organic material that would otherwise shield microorganisms from sterilisation. Practices use:
- Washer-disinfectors, the preferred method, which give validated, consistent and safer results.
- Ultrasonic cleaners, which loosen debris from hinges and serrations and need regular performance testing.
- Manual cleaning, used only where necessary, with heavy-duty gloves, a long-handled brush and the instrument held under water to avoid splashes.
Handpieces have their own cleaning and lubrication routine, often in a dedicated automated unit.
3. Inspection
Under good light, and ideally with magnification, check every instrument for remaining debris, corrosion, damage and blunt edges. Anything still dirty goes back through cleaning. Damaged instruments are removed from service.
4. Packaging
Instruments that must be sterile at the point of use are packed in sterilisation pouches or wrapped cassettes, with a chemical indicator. Pouches are sealed correctly, labelled with the date and steriliser load or cycle number, and never overfilled.
5. Sterilisation
Most dental practices use a benchtop steam steriliser with a cycle suited to wrapped and hollow items, often called a Type B or Type S cycle. Correct loading matters: do not stack pouches, place them paper-to-plastic as the manufacturer recommends, and let steam reach every surface.
6. Monitoring and release
| Check | What it tells you | When |
|---|---|---|
| Physical parameters (printout or data logger) | Time, temperature and pressure were reached | Every cycle |
| Chemical indicators | The pack was exposed to the right conditions | Every pack or load |
| Air-removal / steam penetration tests | Steam can reach inside hollow loads | Daily or as the manufacturer specifies |
| Annual validation | The steriliser performs as intended with your typical loads | At least yearly, and after major repairs |
Only when every check passes is the load released and recorded. A failed cycle means the whole load is reprocessed and the cause investigated.
7. Storage
Allow packs to dry fully before handling. Store sterile items in a clean, dry, enclosed area away from splashes and aerosols, rotating stock so older packs are used first. A pack that is torn, wet or dropped on a contaminated surface is no longer sterile.
8. Traceability
Practices must be able to link a sterilised pack to the patient it was used on. Typically, the pack label (date and cycle) is recorded in the patient's notes or scanned into the practice software. If a cycle is later found to have failed, the affected patients can be identified.
Common mistakes we see
- overloading the steriliser to save time
- skipping inspection because the instrument came out of the washer
- handling packs while still wet
- forgetting to record cycle numbers in patient notes
- mixing clean and dirty workflows in a cramped reprocessing room
Our infection control course covers reprocessing in detail and includes hands-on practice. For the wider picture, read infection control in dental practice.
Frequently asked questions
What replaced AS/NZS 4815?
AS 5369, the Australian Standard for reprocessing reusable medical devices and other devices, now applies to dental practices along with the Dental Board's infection control guidelines.
Do all dental instruments need to be packaged before sterilising?
Instruments that must be sterile when used, such as those for surgical procedures, must be packaged. Requirements for other items depend on how they are used and stored, so follow your practice protocol and the current guidelines.